Table 3 Relative reference normalised errors between the ionization measurements (PTW 31010) and the TPS calculated dose in the CIRS 002 LFC thorax phantom for 8 clinical cases

From: MR compatible detectors assessment for a 0.35 T MR-linac commissioning

Clinical case

No. 1

No. 2

No. 3

No. 4

No. 5

No. 6

No. 7

No. 8

Relative reference normalised error (%)

Point 1

0.9%

 − 0.4%

      

Point 2

    

0.1%

   

Point 5

0.3%

 

0.2%

 − 0.1%

 

 − 1.7%

 − 0.3%

 − 1.0%

Point 7

0.3%

  

0.4%

 − 1.7%

 − 1.3%

  

Point 8

   

 − 0.3%

    

Point 10

 − 0.5%

  

0.8%

 

0.9%