Fig. 5 | Radiation Oncology

Fig. 5

From: Intrafraction organ movement in adaptive MR-guided radiotherapy of abdominal lesions – dosimetric impact and how to detect its extent in advance

Fig. 5

Maximum RoMSim value over simulation duration for each patient. The duration of evaluated simulation sessions was between 1.6 and 28.7 min and maximum RoMSim values between 1.5 mm and 10.5 mm were determined. No significant correlation was found between RoMSim values and the duration of simulation sessions (p > 0.05)