Table 1 The details of the review items

From: Developing an automatic treatment record review system for quality assurance of patient treatment delivery in radiation therapy

Categories

Details

Parameter consistency

Consistency between treatment delivery parameters and treatment plan parameters, which included scrutiny of machine specifications, modality, energy, beam type, source-to-surface distance (SSD), segment count, monitor units (MU), gantry angle, collimator settings, couch angle, jaw configuration, and multi-leaf collimator (MLC) positions, etc

Treatment completeness

Delivery of all treatment fields in the plan

Delivered MU did not exceed planned MU

Treatment progression

Cumulative dose and remaining dose

Accuracy of the remaining session

Consistency between the daily treatment dose and the prescribed dose

Treatment calendar that has been postponed for an extended period or discontinued altogether

Dose verification prior to stereotactic treatment

Image guidance

Image approval in accordance with departmental policies

Applied shifts

Isocenter on the CBCT(cone-beam computed tomography) matched plan

Selection of IGRT (image-guided radiation therapy) scan template and parameters meet clinical requirements

IGRT frequency adherence to medical directives

IGRT registration deviation did not exceed predefined threshold

Overrides

Override records, including instances such as couch position exceeding tolerance, inconsistent field parameters, abnormal dose tracking, treatment fractionation mode not consistent with the prescription, and any other deviations

Treatment couch

Discrepancies between the treatment couch position (vertical, lateral, longitudinal, and rotational) and the reference couch position

Documentation

Completeness of treatment-related documentation

Approval of documents by both the planning physicist and review physicist

Treatment mode

Treatments completed out of clinical mode

The individual performing the QA model treatment