From: Gantry-based pencil beam scanning proton therapy for uveal melanoma: IMPT versus proton arc therapy
Patient | Diseased eye /Prior RT | Tumor | Gantry angle-Couch angle-Range shifter (cm) | ||
---|---|---|---|---|---|
Volume [cc] | Diameter/Thickness [cm] | IMPT | Arc | ||
1 | Left/No | 1.09 | 1.6/0.6 | G325C0R0 G115C0R0 G135C0R0 | Arc plan parameters are valid for all patients and only depend on the disease side (left/right). Left eye: G: 85–135 C: 30/330 R: 0 Right eye: G: 225–275 C: 30/330 R: 0 |
2 | Right/Right | 0.71 | 1.5/0.5 | G30C0R0 G270C0R0 G270C90R0 G240C0R0 | |
3 | Right/Right | 0.77 | 1.5/0.2 | G280C0R2 G270C30R2 G260C0R2 G240C0R2 | |
4 | Left/No | 0.30 | 1.0/0.3 | G90C0R2 G125C0R2 G90C335R2 | |
5 | Right/No | 0.34 | 0.8/0.1 | G300C0R3 G270C0R3 G240C0R3 | |
6 | Right/No | 2.98 | 2.5/0.4 | G270C0R0 G250C0R0 G230C0R0 G230C40R0 | |
7 | Left/No | 0.30 | 0.8/0.5 | G120C0R2 G90C330R2 G85C0R2 | |
8 | Left/No | 0.24 | 0.5/0.4 | G120C0R2 G90C0R2 G90C330R2 | |
9 | Right/No | 0.42 | 1.0/0.6 | G270C30R2 G265C0R2 G240C0R2 | |
10 | Left/No | 1.47 | 2.0/0.4 | G125C0R2 G100C330R2 G250C90R0 | |
11 | Right/No | 0.38 | 1.0/0.4 | G260C0R2 G250C0R2 G235C0R0 |